A '''pH Meter''' is not only considered by some to be an essential tool for all grain homebrewing and commercial brewing, but is also a very useful tool to have for making sour beer.  For examples of when an accurate pH reading is useful in the sour brewing process, see the [[Mixed Fermentation]], [[Sour Worting]], and [[Lactobacillus]] pages.   
+
A '''pH Meter''' is not only considered by some to be an essential tool for all grain homebrewing and commercial brewing but is also a very useful tool to have for making sour beer.  For examples of when an accurate pH reading is useful in the sour brewing process, see the [[Mixed Fermentation]], [[Wort Souring]], and [[Lactobacillus]] pages.   
  
 
==Recommend pH Meters==
 
==Recommend pH Meters==
  
 
===High End pH Meters===
 
===High End pH Meters===
Higher end pH meters recommended by our members and others.
+
Higher end pH meters recommended by our members and others.  "Benchtop" meters are generally higher end, and are good choices for professional brewers.
  
 
* [http://www.milwaukeeinst.com/site/products/products/martini-bench-meters/67-products-g-martini-bench-meters-g-mi150 Mi150 pH and temperature Laboratory Bench Meter] (runs about $250 US) <ref>[https://www.facebook.com/groups/MilkTheFunk/permalink/1058773210817559/?comment_id=1059095470785333&offset=0&total_comments=5&comment_tracking=%7B%22tn%22%3A%22R0%22%7D Conversation with Jonathan Newman from The Virginia Beer Company on Milk The Funk.  04/19/2015.]</ref>
 
* [http://www.milwaukeeinst.com/site/products/products/martini-bench-meters/67-products-g-martini-bench-meters-g-mi150 Mi150 pH and temperature Laboratory Bench Meter] (runs about $250 US) <ref>[https://www.facebook.com/groups/MilkTheFunk/permalink/1058773210817559/?comment_id=1059095470785333&offset=0&total_comments=5&comment_tracking=%7B%22tn%22%3A%22R0%22%7D Conversation with Jonathan Newman from The Virginia Beer Company on Milk The Funk.  04/19/2015.]</ref>
 
* [http://www.denverinstrument.com/denverusa/electrochemistry/UltraBasic_benchtop_meters.php UltraBasic Benchtop Meters] (discontinued, but may find used at a good price) <ref>[https://www.facebook.com/groups/MilkTheFunk/permalink/1058773210817559/?comment_id=1060334447328102&offset=0&total_comments=9&comment_tracking=%7B%22tn%22%3A%22R1%22%7D Conversation with Brandon Jones of Embrace The Funk on Milk The Funk.  04/21/2015.]</ref>.
 
* [http://www.denverinstrument.com/denverusa/electrochemistry/UltraBasic_benchtop_meters.php UltraBasic Benchtop Meters] (discontinued, but may find used at a good price) <ref>[https://www.facebook.com/groups/MilkTheFunk/permalink/1058773210817559/?comment_id=1060334447328102&offset=0&total_comments=9&comment_tracking=%7B%22tn%22%3A%22R1%22%7D Conversation with Brandon Jones of Embrace The Funk on Milk The Funk.  04/21/2015.]</ref>.
 +
* [https://www.coleparmer.com/i/oakton-ph-6-handheld-meter-with-atc-probe-order-ph-probe-separately/3561320?PubID=UX&persist=true&ip=no&gclid=COjPoKPfiNMCFUpNfgodxOsLRA Oakton pH 6+ Handheld Meter with ATC Probe <ref>[https://docs.google.com/document/d/1IHYluAgIFvXYDHMO2WRYzwSSvNo8UtmT4hRYAA_DGpM/edit?usp=sharing Ryan Buxton.  Milk The Funk Facebook group.  04/03/2017.]</ref>.]
 +
* [https://www.thermofisher.com/order/catalog/product/STARA1210 Orion™ Star A121] with an [https://www.thermofisher.com/order/catalog/product/8156BNUWP Orion™ 8156BNUWP ROSS Ultra™], recommended by [https://www.facebook.com/groups/MilkTheFunk/permalink/2044032758958261/?comment_id=2044272178934319&comment_tracking=%7B%22tn%22%3A%22R%22%7D Shawn Savuto] (does not include ATC in this probe).
  
 
===Budget pH Meters===
 
===Budget pH Meters===
Line 27: Line 29:
 
''Disadvantages'':
 
''Disadvantages'':
 
* Despite having an automatic calibration message, this pH meter still often needs to be calibrated before each brew day <ref name="ajdelange"></ref>.
 
* Despite having an automatic calibration message, this pH meter still often needs to be calibrated before each brew day <ref name="ajdelange"></ref>.
* Interface isn't intuitive, and the buttons have been reported to be touchy by some.  Use [http://www.hach.com/asset-get.download.jsa?id=18952100085 the website manual] instead of the instructions that come with it.
+
* The buttons have been reported to be touchy by some
 +
* Poor instructions that come with the unit.  Use [http://www.hach.com/asset-get.download.jsa?id=18952100085 the website manual] instead of the instructions that come with it.
  
====[http://www.milwaukeetesters.com/MW102.html Milwaukee MW102 pH/Temp Meter (updated/upgraded from SM101 and MW101)].====
+
====[http://www.milwaukeeinstruments.com/site/products/products/standardportable-meters/81-products-g-standard-portable-meters-g-mw102 Milwaukee MW102 pH/Temp Meter (updated/upgraded from SM101 and MW101)].====
 
[[File:MW102.jpg|thumb|Milwaukee MW102]]
 
[[File:MW102.jpg|thumb|Milwaukee MW102]]
 
''Features'':
 
''Features'':
Line 40: Line 43:
 
* 0.02+/- accuracy.
 
* 0.02+/- accuracy.
 
* Automatic temperature compensation.
 
* Automatic temperature compensation.
* [http://www.milwaukeetesters.com/pdf/product-specs/MW100-101-102-500.pdf Product specs comparison chart.]
 
* [http://www.milwaukeeinstruments.com/site/pdf/pH_Application_Info.pdf Milwaukee Instruments pH General Information Sheet]
 
 
''Disadvantages'':
 
''Disadvantages'':
 
* This pH meter usually needs to be calibrated before each brew day.
 
* This pH meter usually needs to be calibrated before each brew day.
 
* Probe replacements appear to be difficult to find.  [http://www.eseasongear.com/miinmadojuph.html Here is one vendor].  eBay has been reported to be another option <ref>[https://www.facebook.com/groups/MilkTheFunk/permalink/1065381353490078/?comment_id=1067202043308009&offset=0&total_comments=32&comment_tracking=%7B%22tn%22%3A%22R3%22%7D Conversation with PJ Dunn on MTF.  05/03/2015.]</ref>.  [http://www.ebay.com/itm/161630172779?_trksid=p2057872.m2749.l2649&ssPageName=STRK%3AMEBIDX%3AIT Here is a seller] that PJ Dunn has had good results from, although he reported slow shipping.
 
* Probe replacements appear to be difficult to find.  [http://www.eseasongear.com/miinmadojuph.html Here is one vendor].  eBay has been reported to be another option <ref>[https://www.facebook.com/groups/MilkTheFunk/permalink/1065381353490078/?comment_id=1067202043308009&offset=0&total_comments=32&comment_tracking=%7B%22tn%22%3A%22R3%22%7D Conversation with PJ Dunn on MTF.  05/03/2015.]</ref>.  [http://www.ebay.com/itm/161630172779?_trksid=p2057872.m2749.l2649&ssPageName=STRK%3AMEBIDX%3AIT Here is a seller] that PJ Dunn has had good results from, although he reported slow shipping.
* Probe has to be stored in a storage solution.
+
* Probe has to be stored in [http://www.amazon.com/Milwaukee-MA9015-Storage-Solution-Electrodes/dp/B002P35Y6U Milwaukee MA9015 Storage Solution].
  
 
==Buffer Solution==
 
==Buffer Solution==
Buffer solution will be required to calibrate a pH meter.  In general, it is recommended that a pH meter have a 2 point calibration method, which requires 4.0 buffer solution and 7.0 buffer solution.  Many buffer solutions for pH meter calibration can be found on [http://www.amazon.com/s/?ie=UTF8&keywords=ph+meter+calibration+solution&tag=googhydr-20&index=aps&hvadid=33870947755&hvpos=1t2&hvexid=&hvnetw=g&hvrand=2358411058820273529&hvpone=&hvptwo=&hvqmt=b&hvdev=c&ref=pd_sl_9h5a0pkyc9_b Amazon] or other retailers.  Note that these buffer solutions do have expiration dates, and using expired calibration solutions can throw off pH meter readings.
+
Buffer solution will be required to calibrate a pH meter, and generally has to be purchased separately from the pH meter itself.  In general, it is recommended that a pH meter have a 2 point calibration method, which requires 4.0 buffer solution and 7.0 buffer solution.  Many buffer solutions for pH meter calibration can be found on [http://www.amazon.com/s/?ie=UTF8&keywords=ph+meter+calibration+solution&tag=googhydr-20&index=aps&hvadid=33870947755&hvpos=1t2&hvexid=&hvnetw=g&hvrand=2358411058820273529&hvpone=&hvptwo=&hvqmt=b&hvdev=c&ref=pd_sl_9h5a0pkyc9_b Amazon] or other retailers.  Note that these buffer solutions do have expiration dates, and using expired calibration solutions can throw off pH meter readings.
  
 
==Testing the Probe==
 
==Testing the Probe==
 
The probes on pH meters generally have a limited life, depending on the quality of the unit.  Milwakuee Instruments describes the following method for testing the life of the probe:
 
The probes on pH meters generally have a limited life, depending on the quality of the unit.  Milwakuee Instruments describes the following method for testing the life of the probe:
  
''"A simple test can be performed using Windex glass cleaner with ammonia and white distilled vinegar. The procedure is as follows: Turn your meter or tester on and place the probe in white distilled vinegar which is acidic (2.4ph) and the reading will be from 2.3 to 2.5 but must be below 3.0 then go to the Windex with ammonia which is alkaline. Your display should move very quickly up the scale to 10.5 – 11.2 but must go above 10.0.  If the probe slowly moves up the scale then it is time to consider replacing it. If it does not go below 3.0 pH in vinegar and/or above the 10.0 pH in Windex then probe is dead."'' <ref>[http://www.milwaukeeinstruments.com/site/pdf/pH_Application_Info.pdf Milwaukee Instruments pH General Information Sheet.  Retrieved 4/30/2015.]</ref>
+
"A simple test can be performed using Windex glass cleaner with ammonia and white distilled vinegar. The procedure is as follows: Turn your meter or tester on and place the probe in white distilled vinegar which is acidic (2.4ph) and the reading will be from 2.3 to 2.5 but must be below 3.0 then go to the Windex with ammonia which is alkaline. Your display should move very quickly up the scale to 10.5 – 11.2 but must go above 10.0.  If the probe slowly moves up the scale then it is time to consider replacing it. If it does not go below 3.0 pH in vinegar and/or above the 10.0 pH in Windex then probe is dead." <ref>[http://www.milwaukeeinstruments.com/site/pdf/pH_Application_Info.pdf Milwaukee Instruments pH General Information Sheet.  Retrieved 4/30/2015.]</ref>
 +
 
 +
==Using a pH Meter==
 +
(This section in progress)
 +
 
 +
* [http://accidentalis.com/measuring-mash-ph-in-practice/ "Measuring Mash pH in Practice," by Matt Crispen (a good general guide to using a pH meter).]
 +
* [https://www.morebeer.com/articles/understanding_ph_in_brewing "Understanding pH and It's Application in Small-Scale Brewing," by AJ Delange.]
 +
* Carbonic acid can affect pH readings.  If testing a beer that has CO2 in it, either residual from fermentation or from keg/bottle carbonation, let the beer warm to room temperature and swirl the beer to try and get all of the CO2 out of solution before taking a pH measumrement <ref>[https://www.facebook.com/groups/MilkTheFunk/permalink/1153226544705558/?comment_id=1153275738033972&reply_comment_id=1153651057996440&total_comments=5&comment_tracking=%7B%22tn%22%3A%22R%22%7D Conversation with Gareth Young on MTF.  09/23/2015.]</ref>.
  
 
==External Articles==
 
==External Articles==
Line 61: Line 69:
 
* [https://www.facebook.com/permalink.php?story_fbid=851663348222683&id=464551136933908 Bru'n Water on how to use a pH meter.]
 
* [https://www.facebook.com/permalink.php?story_fbid=851663348222683&id=464551136933908 Bru'n Water on how to use a pH meter.]
 
* [http://embracethefunk.com/ph-readings-of-commercial-beers/ pH Readings of Commercial Beers, Embrace the Funk Blog, Brandon Jones.]
 
* [http://embracethefunk.com/ph-readings-of-commercial-beers/ pH Readings of Commercial Beers, Embrace the Funk Blog, Brandon Jones.]
 +
* [http://bitesizebio.com/8750/how-to-care-for-your-ph-meter/ How to Care for Your pH Meter; BitesizeBio.]
 +
* [https://sites.google.com/site/brunwater/water-knowledge "Water Knowledge" by Martin Brungard.]
  
 
==References==
 
==References==

Revision as of 20:43, 24 May 2022

A pH Meter is not only considered by some to be an essential tool for all grain homebrewing and commercial brewing but is also a very useful tool to have for making sour beer. For examples of when an accurate pH reading is useful in the sour brewing process, see the Mixed Fermentation, Wort Souring, and Lactobacillus pages.

Recommend pH Meters

High End pH Meters

Higher end pH meters recommended by our members and others. "Benchtop" meters are generally higher end, and are good choices for professional brewers.

Budget pH Meters

pH meters can range from $30 to $500+. The problem with many of the cheaper pH meters is that their reliability and longevity are often questionable [4]. Commonly, two pH meters are recommended by trustworthy sources (AJ Delange and Kai Troester).

Hach Pocket Pro+ pH Tester with Replaceable Sensor.

Hach Pocket Pro+

Features:

  • This pH meter is recommended by AJ Delange [4][5].
  • Does not require storage solution (there is an o-ring seal on the cap, which only needs to have a few drops of distilled water in it).
  • Replaceable sensor.
  • Stability [4].
  • 2 Buffer calibration.
  • Electrode longevity [4].
  • Junction resistant to fouling by sugars, proteins [4].
  • Lets the user decide when to accept a buffer calibration reading [4].
  • ATC [4].
  • Automatic buffer recognition [4].

Disadvantages:

  • Despite having an automatic calibration message, this pH meter still often needs to be calibrated before each brew day [4].
  • The buttons have been reported to be touchy by some.
  • Poor instructions that come with the unit. Use the website manual instead of the instructions that come with it.

Milwaukee MW102 pH/Temp Meter (updated/upgraded from SM101 and MW101).

Milwaukee MW102

Features:

  • This pH meter is recommended by Kai Troester (SM101 originally) [6].
  • 2 Buffer calibration (calibration video)
  • ATC.
  • Stability [7].
  • Separate pH and temperature probes (can use as RTD for temp monitoring) [8].
  • Storage solution container screws into probe [8].
  • 0.02+/- accuracy.
  • Automatic temperature compensation.

Disadvantages:

Buffer Solution

Buffer solution will be required to calibrate a pH meter, and generally has to be purchased separately from the pH meter itself. In general, it is recommended that a pH meter have a 2 point calibration method, which requires 4.0 buffer solution and 7.0 buffer solution. Many buffer solutions for pH meter calibration can be found on Amazon or other retailers. Note that these buffer solutions do have expiration dates, and using expired calibration solutions can throw off pH meter readings.

Testing the Probe

The probes on pH meters generally have a limited life, depending on the quality of the unit. Milwakuee Instruments describes the following method for testing the life of the probe:

"A simple test can be performed using Windex glass cleaner with ammonia and white distilled vinegar. The procedure is as follows: Turn your meter or tester on and place the probe in white distilled vinegar which is acidic (2.4ph) and the reading will be from 2.3 to 2.5 but must be below 3.0 then go to the Windex with ammonia which is alkaline. Your display should move very quickly up the scale to 10.5 – 11.2 but must go above 10.0. If the probe slowly moves up the scale then it is time to consider replacing it. If it does not go below 3.0 pH in vinegar and/or above the 10.0 pH in Windex then probe is dead." [10]

Using a pH Meter

(This section in progress)

External Articles

References